Mizota Tadato
Affiliate Master
Yamaguchi University
An Improvement of a Twin Conduction-type Microcalorimeter and the Data Processing System
Memoirs of the Faculty of Engineering, Yamaguchi University Volume 32 Issue 1
Page 57-62
published_at 1981
Title
双子伝導型微少熱量計の改良とデータ処理システムの作成
An Improvement of a Twin Conduction-type Microcalorimeter and the Data Processing System
Source Identifiers
A twin conduction-type microcalorimeter (Rhesca CM-204-D1) has been improved especially on its agitation mechanism and ampoule-breaker. These mechanisms were simplified that the duration up to thermal equilibrium was shortened and the base line was stabilized. A simple digital data processing system controlled by the microcomputer MZ80K has been made for the data acquisition and calculation required. Heats of immersion of both rutile and anatase in water have been obtained as a function of sample weight and the results were 350 and 500 erg/(cm)^2 respectively, for the samples outgassed at 400℃ for 1 h in vacuo. Lower limit for heat-detection of this calorimeter system under conditions applied to the present case was 0.1-0.2 joule on the heat of immersion experiment.
Languages
jpn
Resource Type
departmental bulletin paper
Publishers
山口大学工学部
Date Issued
1981
File Version
Version of Record
Access Rights
open access
Relations
[ISSN]0372-7661
[NCID]AN00244228
Schools
工学部