Measurement and imaging of cerebral blood flow with stable xenon and computed tomography using modified least-squares analysis
山口医学 Volume 37 Issue 4
Page 247-253
published_at 1988-08
Title
Modified least-squares法を用いた非放射性キセノン吸入によるコンピューター断層脳血流検査
Measurement and imaging of cerebral blood flow with stable xenon and computed tomography using modified least-squares analysis
Creators
Yamashita Tetsuo
Creators
Nakano Shigeki
Creators
Hayashi Masaaki
Creators
Abiko Seisho
Creators
Aoki Hideo
Source Identifiers
Studies of stable xenon CT CBF technique with a newly developed closed system inhalator are reported. The patient inhales 30% of xenon for six minutes using a closed system(XETRON 3) while a series of CT scans is obtained using a Somaton DR 3 CT scanner. Endtidal xenon concentration is measured using a xenon monitor (AZ-723-XS), and CBF is calculated with Siemens medical CT CBF imaging software. The software uses Koeppe's mocified nonlinear least-squares analysis for calculation of f and k in Kety-Schmidt's equation. The scan slice level was selected about 5 cm above and parallel to orbitomeatal line, showing the pineal body and both glomus of the choroid plexus. CBF was evaluated in the region of interest corresponding to anatomical structure. Ten cases were studied including five normal adults, three cases of headache, one of small acoustic neurinoma and one of hemifacial spasm. There were seven males and three females, with ages ranging from 25 to 59 years old and averaging 43. In this system, normal averaging CBF (ml/100g/min) of left and right hemisphere was 52.0±9.0 and 51.1±7.4, respectively. The results are reasonable and this technique has the potential for broad utilization as a cheap, less-invasive three dimensional method of clinical examination for the evaluation of brain disorders.
Languages
jpn
Resource Type
journal article
Publishers
山口大学医学会
Date Issued
1988-08
File Version
Not Applicable (or Unknown)
Access Rights
metadata only access
Relations
[ISSN]0513-1731
[NCID]AN00243156
Schools
医学部