Quantification of serum seleium by a polarized Zeeman atomic absorption spectrophotometer

山口医学 Volume 39 Issue 5 Page 601-605
published_at 1990-10
Title
ゼーマン偏光原子吸光装置による血清セレンの測定
Quantification of serum seleium by a polarized Zeeman atomic absorption spectrophotometer
Creators Kimura Takeshi
Creators Sugino Kayoko
Creators Tamura Yoichi
Creators Furunaga Akihiko
Source Identifiers
Selenium concentration in blood serum was determined by a polarized Zeeman atomic absortion sepctrophotometer. Ten μl of a sample solution (5×dilution of serum, with 0.1g/dl rhodium nitrate and 0.25% Triton X 100) was pipetted into a pyrolytic carbon-coated graphite cuvette. Between-day coefficient of varriation for the means of quadruple assays on a normal serum (selenium 7.17μg/dl) was 7.7% (n=10). Normal calues for adults (24 males and 26 females) were 8.40±1/92μg/dl (mean±standard deviation).