Development of a palpation sensor for detecting prostatic cancer and hypertrophy : signal processing on clinical data

Transactions of the Japan Society of Mechanical Engineers. C Volume 65 Issue 636 Page 3296-3301
published_at 1999-08
2010010181.pdf
[fulltext] 503 KB
Title
前立腺の硬さ及び形状評価のための触診センサ : 臨床応用と信号処理について
Development of a palpation sensor for detecting prostatic cancer and hypertrophy : signal processing on clinical data
Creators Tanaka Mami
Creators Kamei Masatomo
Creators Furubayashi Mitsuyuki
Creators Jiang Zhong Wei
Creators Tanahashi Yoshikatu
Creators Chonan Seiji
Creator Keywords
palpation sensor PVDF film measurement of organic stiffness prostatic cancer and hypertrophy clinical test
This paper is concerned with the development of a palpation sensor for possible detection of prostatic cancer and hypertrophy. The sensor is a layered medium with two polyvinylidene fluoride (PVDF) films embedded in the matrix sponge rubber. This sensor fixed with a finger is pressed against the prostates and the output voltage signals from the polymer films are measured. The temporal mean values, I_1 and I_2, on the absolute signals from polymer films are calculated and their ratio I_1/I_2 is used as the sensor output. First, the experimental method for the clinical test is proposed. The prostate glands diagnosed by a doctor's fore finger are classified into four groups according to their stiffness and the data processing for assorting the glands into the groupes are presented. The effect of the convexes placed on the surface of sensor is examined by conparing the clinical data with that of the sensor without the convexes.
Languages jpn
Resource Type journal article
Publishers 日本機械学会
Date Issued 1999-08
Rights
Copyright: 日本機械学会 (The Japan Society of Mechanical Engineers)()
File Version Version of Record
Access Rights open access
Relations
[ISSN]0387-5024
[NCID]AN00187463
[isVersionOf] [URI]http://www.jsme.or.jp/publish/ronbun/
Schools 大学院理工学研究科(工学)